Digital Clamp Meters Recalled by Fluke Due to Shock Hazard
Burn
The hazard
The meters can fail to give an appropriate voltage reading, resulting in the operator falsely believing the electrical power is off, posing a shock, electrocution, or thermal burn hazard.
Incidents reported: Fluke has received three reports of clamp meters displaying an incorrect voltage reading. No injuries have been reported.
What to do now
Consumers should stop using these recalled Digital Clamp Meters immediately and contact Fluke for a free replacement clamp meter.
Contact: For additional information, contact Fluke toll-free at (888) 983-5853 between 7 a.m. and 4 p.m. PT Monday through Friday or visit the firm's website at www.fluke.com/33Xrecall
About this recall
This recall involves Fluke Digital Clamp Meters with model and serial numbers listed below. The model number may be followed by an "A." Model No. Starting With Serial No. Starting with Serial No. Ending In 333 97010000 98880240 334 97010000 98870101 335 97010000 98860104 336 96220000 98900400 337 96070000 98890607 Serial numbers that begin with "S" or are followed by "R" are not included in the recall. "Fluke" and the model number are printed on the front of the unit and the serial number on the back. The instrument body is yellow, red, and black. The clamp meters measure 0 to 600 volts alternating current (VAC), 0 to 600 volts direct current (VDC) and 0 to 400, 600 or 1000 amps alternating current.
Product photos
Photos: U.S. CPSC